제품 상세보기 SINOPRO MAX

Home   >   제품 상세보기

제품 상세보기

Product
Eight-channel battery test CT-4008T-5V50mA-164
Product No.
CT-4008T-5V50mA-164
Brand
SINOPROMRX

DESCRIPTION

Product name: Battery testing system CT-4008T-5V50mA-164

SPECIFICATION

Input AC

AC 220V/110V ±10% / 50Hz

Input power

25 W

Resolution

AD:16bit; DA:16bit

Input impedance

≥1GΩ

Voltage

CV output range

25mV~5V

Min discharge voltage

-5V

Accuracy

± 0.05% FS

Stability

± 0.05% FS

Current

Output range/channel

Range1: 5μA-1mA; Range2: 1mA-25mA; Range3: 25mA-50mA

Accuracy

± 0.05% FS

CV cut-off current

Range1: 2μA; Range2: 50μA; Range3: 100μA

Stability

± 0.05% FS

Power

Output power/channel

0.25W

Stability

± 0.1% FS

Time

Current response time

<500μs (10% FS~90% FS)

Working step time

≤(365*24)h/step Time format-00:00:00.000(h, m, s, ms)

Data record

 

 

Data record conditions

Min data record interval:100ms

Min voltage change: 10mV

Min current change: 2μA; 50μA; 100μA

Frequency

10Hz

Charge

Charge modes

CC, CV, CCCV, CP

Cut-off condition

Voltage, Current, Time, Capacity, Energy, -△V

Discharge

Discharge modes

CC, CV, CCCV, CP, CR

Cut-off condition

Voltage, Current, Time, Capacity, Energy

Pulse

Charge

CC, CP

Discharge

CC, CP

Min pulse width

500ms

Pulse counts

Up to 32

Chg and Dschg switch

Supported

Cut-off condition

Voltage, Time

DCIR

Calculated DCIR values available in BTSDA

Cycle

Max cycles

65535

Max steps

254

Cycle nest

3

Protection

 

 

 

Safety protection

Power-failure data protection

Offline operation mode

User-defined protection conditions, such as upper and lower limited current/voltage, delay time, temperature, etc.

Channels feature

Independent pairs of closed loop for constant

Channels control mode

Independent control

Data acquisition method

Kelvin connection

Noise

<45dB

Database

MySQL

Communication

TCP/IP

Operating system

Windows 7/8/10 64bit

Data export

EXCEL, TXT, CSV, PDF, Plot/Graph

Min. disk drive

500GB

Communication port

Ethernet port

Leak Current

<0.005μA

Channels per unit

8

Operation and storage environment requirement

Operation environment temperature

0℃~40℃ (When the temperature is 25±10℃, the accuracy offset caused by temperature is less than 50ppm/℃)

Storage environment temperature

-10℃~50℃

Operation environment humidity

≤70% RH(no moisture condensation)

Storage environment humidity

≤80% RH(no moisture condensation)

Clamps and dimensions

Tester pictures

Clamps pictures

Universal holder

Alligator

Polymer

Coin

Pictures just for reference

Dimensions

(W*D*H)(mm)

1U19", 483*310*48

Warranty

One year limited warranty with lifetime support