Home > 제품 상세보기
DESCRIPTION
Product name: Battery testing system CT-4008T-5V6A-S1
SPECIFICATION
Input AC |
AC 220V/110V ±10% / 50Hz |
|||
Input power |
425 W |
|||
Resolution |
AD:16bit; DA:16bit |
|||
Input impedance |
≥1MΩ |
|||
Voltage |
CV output range |
25mV~5V |
||
Min discharge voltage |
1V for universal holder,1.5V for 2m cable |
|||
Accuracy |
± 0.05% FS |
|||
Stability |
± 0.05% FS |
|||
Current |
Output range/channel |
Range1: 0.5mA-0.1A; Range2: 0.1A-3A; Range3: 3A-6A |
||
Accuracy |
± 0.05% FS |
|||
CV cut-off current |
Range1: 0.2mA; Range2: 6mA; Range3: 12mA |
|||
Stability |
± 0.05% FS |
|||
Power |
Output power/channel |
30W |
||
Stability |
± 0.1% FS |
|||
Time |
Current response time |
<1ms (10% FS~90% FS) |
||
Working step time |
≤(365*24)h/step Timeformat-00:00:00.000(h, m, s, ms) |
|||
Data record |
Data record conditions |
Min data record interval:100ms |
||
Min voltage change: 10mV |
||||
Min current change: 0.2mA; 6mA; 12mA |
||||
Frequency |
10Hz |
|||
Charge |
Charge modes |
CC, CV, CCCV, CP |
||
Cut-off condition |
Voltage, Current, Time, Capacity, Energy, -△V |
|||
Discharge |
Discharge modes |
CC, CV, CCCV, CP, CR |
||
Cut-off condition |
Voltage, Current, Time, Capacity, Energy |
|||
Pulse |
Charge |
CC, CP |
||
Discharge |
CC, CP |
|||
Min pulse width |
500ms |
|||
Pulse counts |
Up to 32 |
|||
Chg and Dschg switch |
Supported |
|||
Cut-off condition |
Voltage, Time |
|||
DCIR |
Calculated DCIR values available in BTSDA |
|||
Cycle |
Max cycles |
65535 |
||
Max steps |
254 |
|||
Cycle nest |
3 |
|||
Protection |
Safety protection |
Power-failure data protection |
||
Offline operation mode |
||||
User-defined protection conditions, such as upper and lower current/voltage, delay time, temperature, etc. |
||||
Channels feature |
Independent pairs of closed loop for constant |
|||
Channels control mode |
Independent control |
|||
Data acquisition method |
Kelvin connection |
|||
Noise |
<55dB |
|||
Database |
MySQL |
|||
Communication |
TCP/IP |
|||
Operating system |
Windows 7/8/10 64bit |
|||
Data export |
EXCEL, TXT, CSV, PDF, Plot/Graph |
|||
Min. disk drive |
500GB |
|||
Communication port |
Ethernet port |
|||
Leak Current |
<5μA |
|||
Channels per unit |
8 |
|||
Operation and storage environment requirement |
||||
Operation environment temperature |
0℃~40℃ (When the temperature is 25±10℃, the accuracy offset caused by temperature is less than 50ppm/℃) |
|||
Storage environment temperature |
-10℃~50℃ |
|||
Operation environment humidity |
≤70% RH(no moisture condensation) |
|||
Storage environment humidity |
≤80% RH(no moisture condensation) |
|||
Clamps and dimensions |
||||
Tester pictures |
|
|||
Clamps pictures |
|
|
|
|
Universal holder |
Alligator |
Polymer |
Ring connector |
|
Pictures just for reference |
||||
Dimensions (W*D*H)(mm) |
3U19", 483*404*130 |
|||
Warranty |
One year limited warranty with lifetime support |