Home > 제품 상세보기
DESCRIPTION
Product name: Internal resistance tester MRX-BT
SPECIFICATION
Measurement methods |
Four terminal test method |
||
Voltage |
Voltage measurement range |
0V~20V |
|
Precision |
± (0.1% of FS + 0.1% of reading) |
||
resistance |
Resistance measurement range |
200mΩ2000mΩ two kinds can be automatically switched |
|
Precision |
0mΩ ~200mΩ ± (0.5% of FS + 0.2% of reading) |
||
200mΩ~2000mΩ ± (0.5% of FS + 0.2% of reading) |
|||
A/D method |
12-bit SAR successive comparison method |
||
frequency |
Test frequency |
1kHz |
|
Sampling Rate |
100 times/S |
||
Display refresh rate |
5 times/S |
||
Test current |
MAX AC 20mA |
||
Comparators |
Open method |
Manually open/close |
|
Quantity |
10 sets of voltage and internal resistance upper and lower limits |
||
Abnormal alarm method |
When using the comparator, the buzzer can be set to ON/OFF |
||
When voltage or AC internal resistance exceeds the upper and lower limits of the comparator, the buzzer will sound |
|||
Green indicator light means qualified, red indicator light means unqualified |
|||
Storage function |
Can store up to 100,000 sets of test data |
||
View function |
Can view historical test results |
||
Calibration function |
Zero point calibration |
Calibrate the zero point of voltage and internal resistance at the same time |
|
Voltage calibration |
Use external DC 3V~20V (settable) reference calibration voltage |
||
AC internal resistance calibration |
Use 10~2000mΩ (settable) precision resistance to calibrate internal resistance |
||
External control terminal expansion |
Extensible automated testing, expandable USB communication (optional accessories, not included in the standard configuration) |
||
Preheat time |
More than 3min |
||
Equipment working environment requirements |
|||
range of working temperature |
0℃~40℃ (within the range of 25±10℃, the measurement accuracy is guaranteed: accuracy drift 0.005% of FS / |
||
Storage temperature range |
-10℃~50℃ |
||
Relative humidity range of working environment |
≤70% RH (no condensation) |
||
Storage environment relative humidity range |
≤80% RH (no condensation) |
||
Use place |
Both indoor and outdoor, easy to carry |
||
Fixture specification and equipment size |
|||
annex |
Polymer clamp or special test lead |
||
Charger (5V/1A) |
|||
Standard resistance (100mΩ) |
|||
USB cable |
|||
volume |
W*D*H: 120*75*25(mm) |
||
weight |
About 200g |
||
Warranty |
One year limited warranty with lifetime support |